Electrochemical Analyses
Solid-state electrochemistry is used as a crucial analytical tool to investigate ion and electron transport, decomposition behavior at interfaces and forming interphases, Red-Ox processes of active materials and battery cell performance.
Potentiostats – The variability of potentiostats is crucial for in-depth electrochemical investigations and a multitude of different potentiostats are in use. While the electrochemical cells are mostly custom built, they are typically connected in controlled temperature environments.
Instrument specifications:
| Variable RT application | Standard T-dep. impedance | Variable T-dep. application | Low noise potentiostat application | |
| Voltage range | 10 V | 10 V | 10 V | 10 V | 
| Resolution | 76 µV | 76 µV | 76 µV | 3 µV | 
| Current range | 10 nA to 500 mA | 10 nA to 500 mA | 10 nA to 500 mA | 10 nA to 500 mA * | 
| Resolution | 0.76 pA | 0.76 pA | 0.76 pA | 0.03 pA | 
| Impedance spectroscopy | ||||
| Frequency range | 7MHz - 10µHz | 7MHz - 10µHz | 7MHz - 10µHz | 1MHz - 10µHz | 
| Resolution | <0,0001% | <0,0001% | <0,0001% | <0,003% | 
| VRMS (mV) | 0.1 to 2500 | 0.1 to 2500 | 0.1 to 2500 | 0.2 to 350 | 
| EIS voltage resolution | 0,76 µV for ±25 mV | 0,76 µV for ±25 mV | 0,76 µV for ±25 mV | |
| T-range | 25 °C | −40 to 60°C | −20 to 45°C | 10 to 60 °C | 
*up to 10 A Booster
Under Pressure – Set-ups for pressure monitoring and active pressure control of solid-state cells are available. With pressures up to 1.7 GPa, changes to ionic conductivity can be monitored in-situ in house and have been shown to alter the microstructure of softer solids.
Instrument specifications:
| Custom setup | High pressure | |
| Force uniaxial | up to ~5 kN | 75 kN (up to 1.7 GPa @ 6mm Ø) | 
| Force accuracy | 0.1% | 0.1% | 
| Cells | Ø 10 mm Coin-cell | Ø 6, 12 mm Coin-cell | 
| Temperature | 25 °C | −30 to +250 °C | 
Broadband Impedance spectroscopy – Specialized impedance spectrometers are available for impedance monitoring of (mostly solid) samples set up to be measured between −150 °C up to 1200 °C mostly under inert conditions.
Instrument specifications:
| Low Temperature | High-Temperature | |
| Contacting | 2-wire | 4-wire | 
| Impedance spectroscopy | ||
| Frequency range | 20 MHz to 3 µHz | 20 MHz to 3 µHz | 
| Frequency accuracy | 0.01% | 0.01% | 
| VRMS (mV) | 0.045 to 1360 | 0.045 to 1360 | 
| Temperature Range (non-inert transfer) | −160 to +150°C (+400 °C) | RT to +1050 °C (+1200 °C) | 
| Temperature Resolution | 0.01 °C | 0.1 °C | 
| Heating/cooling rate | 20 °C/min | |
| Modular Potentiostat available | 15 V / 10 A | 15 V / 10 A | 
Battery cycler – Battery cycling capabilities are available at controlled room temperature, or selectively between −20 and 45 °C.
Instrument specifications:
| Voltage range | ± 5V, 6V | 
| Current range | multiple 150 µA up to 5 A | 
| T-range | 25 °C, −40 to 60 °C | 
Interested? – Feel free to request collaborative measurements or instrument access to marvin.kraft@uni-muenster.de or ptill@uni-muenster.de.